Nucleic acid detection box

  • 产品详情


| Size of product


Product name

Specifications

Art.No

Portable multipathogenic nucleic acid detection box

CHIPPCR CASE_A

ZSYQ-01003



| Use the procedure Use Process


微信截图_20240515111808.png

Join the template


Input information


Nucleic acid extraction


On the machine detection


detection result



| Product advantage ADVANTAGES


                                      eae0ae14ec821ba72215758c08b4ed9.png



| Application scenarios


731b96e62e0b12a8ee3568807815884_03.png

731b96e62e0b12a8ee3568807815884_05.png

An on-board laboratory multipathogen screening program



731b96e62e0b12a8ee3568807815884_08.png

核酸检测箱.png

A strong pathogen screening program



731b96e62e0b12a8ee3568807815884_14.png

731b96e62e0b12a8ee3568807815884_16.png

CDC-Rapid screening for infectious diseases



| Performance index



Metric

CHIPPCR CASE_A

Stimulate the wavelength of light

475nm~571nm

Emission light wavelength

525nm~612nm

Outline dimension


Overall size: single channel instrument 500370200mm (length, width and height)

Weight: Single-channel instrument 10.44kg

Working environment parameters


Working environment temperature: 10℃ ~30℃, relative humidity: 10%~80%

Atmospheric pressure: 86.0kPa~106.0kPa

Temperature control module performance


Average heating rate: from 50℃ to 90℃, not less than 6℃ / s;

Maximum heating rate: from 50℃ to 90℃, not less than 6℃ / s.

Average cooling rate: from 90℃ to 50℃, not less than 6℃ / s;

Maximum cooling rate: from 90℃ to 50℃, not less than 6℃ / s.

Temperature uniformity: ± 1℃ intercompartment difference

Temperature accuracy: the absolute value of the difference

between measured value and set temperature is <0.2℃

Storage and transportation of

the environmental conditions

Temperature:-10℃~+55℃

Relative humidity: 95% (no condensation)